Article describing assay update transition
PathoSEEK® Detection Assay Updates
Release Date: August 25, 2025
We’ve updated three assays:
-
Total Aerobic Count (TAC) — PN 420106
-
Bile Tolerant Gram Negative (BTGN) v2 — PN 420149
-
Total Coliform & Entero — PN 420153
What’s New
1. Internal Control
- All three assays now target a new region of the cannabis genome.
- For non-flower matrices, use the PathoSEEK Internal Control (PN 420337)
2. Master Mix Changes
- BTGN and Total Coliform & Entero now use Amplification Mix (instead of qPCR MM v3).
-
Updated Cq → CFU conversion equations apply.
-
- TAC continues to use qPCR MM v3, so its conversion equation has not changed.
3. Assay + Mix Kits
Each updated assay is now sold as a kit that includes the correct Master Mix / Amplification Mix.
4. Automation
qPCR Setup is can be automated on MYRA Liquid Handler and can be run on MIC qPCR Instrument
What has NOT changed
The microbial target sequences
The assay positive controls
Availability
Original versions will remain available until July 2026.
Helpful Website Links:
Total Aerobic Count (TAC) Detection Assay
-
Current Version (v1)
2. Bile Tolerant Gram Negative (BTGN) Detection Assay
-
Current Version (v2)
-
Updated Kit (v3)
3. Total Coliform & Entero Detection Assay
-
Current Version (v1)
-
Updated Kit Version (v2)
Current Assay (PN) |
Updated Kit (PN) |
Key Changes | Notes |
---|---|---|---|
Total Aerobic Count (TAC) PN 420106 |
Total Aerobic Count (TAC) v2 PN 420541 |
• Still uses qPCR MM v3 (no conversion update) • Targets new cannabis genome region • Sold as kit with Master Mix |
Conversion equation unchanged |
Bile Tolerant Gram Negative (BTGN) v2 PN 420149 |
Bile Tolerant Gram Negative (BTGN) v3 PN 420539 |
• Now uses Amplification Mix (instead of qPCR MM v3) • Conversion equations updated • Targets new cannabis genome region • Sold as kit with Amplification Mix |
Internal Control PN 420337 required for non-flower matrices |
Total Coliform & Entero PN 420153 |
Total Coliform & Entero v2 PN 420538 |
• Now uses Amplification Mix (instead of qPCR MM v3) • Conversion equations updated • Targets new cannabis genome region • Sold as kit with Amplification Mix |
Internal Control PN 420337 required for non-flower matrices |